Test methods marked with “*” sign are within the scope of the accreditation.
ISO 7206-4 Determination of endurance properties of stemmed femoral components

*ISO 7206-4 Determination of endurance properties of stemmed femoral components

The endurance properties are determined for hip joint prostheses and femoral components under test condition according to this standard.

ISO 7206-6 Determination of endurance properties of head and neck region of stemmed femoral components

*ISO 7206-6 Determination of endurance properties of head and neck region of stemmed femoral components

According to the standard,  endurance properties of neck region of metal or non metal hip joint prosthesis are determined. Due to the test condition in the standard, the test specimens embedded in a solid material are tested with a cyclic loading which is parallel to neck region of the stem.

According to the standard,  endurance properties of neck region of metal or non metal hip joint prosthesis are determined. Due to the test condition in the standard, the test specimens embedded in a solid material are tested with a cyclic loading which is parallel to neck region of the stem.

ISO 7206-10 Determination of static load reistance of modular femoral head

*ISO 7206-10 Determination of static load reistance of modular femoral head

According to this standard, modular constructions which are used for total or partial hip joint replacement are loaded to failure of femoral head.

ISO 7206-13 Determination of torque resistance of head fixation of stemmed femoral components

*ISO 7206-13 Determination of torque resistance of head fixation of stemmed femoral components

Within the scope of this standard loosen torque of the head of hip joint prosthesis are determined.

ASTM F 1875 Standard Practice for Fretting Corrosion Testing of Modular Implant Interfaces: Hip Femoral Head-Bore and Cone Taper Interface

ASTM F1875 Standard Practice for Fretting Corrosion Testing of Modular Implant Interfaces: Hip Femoral Head-Bore and Cone Taper Interface

For measuring fretting corrosion and determining mechanical stability between head and stem junction of modular hip implant, cyclic loading was subjected under special conditions. In our testing, Method 1 is used for testing, analysis and characterization of specimens.

ASTM F 1820 Standard Test Method for Determining the Forces for disaasembly of acetabular devices

*ASTM F1820 Standard Test Method for Determining the Forces for disaasembly of acetabular devices

According to this standard, measurement of  the attachement strength between acetabular shell and liner is determined.